Dosimetric Evaluation of Different Algorithms on Heterogeneous Slab Phantom Using CMS XiO and MONACO Treatment
Ajay Katake1,2, Lalit Kumar3, Balbir Singh4
1Department of Physics, Lovely Professional University, Phagwara, India.
Aim:
To study the dosimetric behavior of dose computational algorithms in inhomogeneous medium using CMS XiO and MONACO treatment planning system (TPS) for 4 megavoltage (MV), 6 MV and 15 MV photon beam energies.
Material And Methods:
Styrofoam blocks of thickness 1.90 cm, 3.8 cm and 5.70 cm was used to introduce inhomogeneity in a slab phantom. Wipro GE computed tomography (CT) scanner was used to scan the phantom. Doses were computed on the central axis of the beam using convolution (C), superposition (S), fast superposition (FS), collapsed cone convolution (CCC) and monte carlo (MC) algorithms for field geometries of 5x5 cm2 and 10x10 cm2 for above said photon beam energies, respectively. An Ion chamber (IC) of 0.6 cc volume was used for the dose measurements. The deviation between measured and TPS computed doses were recorded.
Results:
The PDD (Percentage depth dose) data obtained from the TPS (calculated data) and LINAC (measured data) was used for comparison based on different algorithms in order to calculate the percentage of maximum deviation (PMD). The PMD in MC algorithm were calculated for field sizes of 5x5 cm2 and 10x10 cm2 are found to be in ranging from 0.73% to -4.49% for 4MV, 1.62% to -2.42% for 6MV and 4.53% to -1.47% for 15 MV for 1.90 cm air gap, 2.21% to -3.75% for 4MV, 3.87% to -2.88% for 6 MV and 4.87% to -3.46% for 15 MV for 3.80 cm air gap, 2.77% to -4.66% for 4MV, 3.87% to -2.86% for 6 MV and 5.66% to -4.92% for 15 MV for 5.70cm air gap which is less as compared to CCC, C, FS, and S algorithms.
Conclusion:
The comparison of C, S, FS, CCC and MC algorithms demonstrated that MC having better agreement with IC measurements. In conclusion, MC is a superior option for dose computation, particularly in the presence of low-density heterogeneities.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...