Optimizing time-of-flight of PET/CT image quality via penalty β value in Bayesian penalized likelihood reconstruction
H Murat1, M A A Zulkifli2, M A Said3
1Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400 Serdang, Selangor, Malaysia; Department of Nuclear Medicine, Hospital Sultanah Aminah, 80100 Johor Bahru, Johor, Malaysia.
Introduction:
Optimizing the image quality of Positron Emission Tomography/Computed Tomography (PET/CT) systems is crucial for effective monitoring, diagnosis, and treatment planning in oncology. This study evaluates the impact of time-of-flight (TOF) on PET/CT performance, focusing on varying penalty β values within Q. Clear reconstruction algorithm.
Methods:
The study measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) using the Discovery MI PET/CT scanner and NEMA IQ phantom filled with the radiotracer fluorodeoxyglucose (18F-FDG). PET/CT scans were performed with and without TOF using β values of 100, 500, 1000, 1500, 2000, and 3000. Pixel intensity values were measured using ImageJ software, and SNR and CNR were calculated.
Results:
Results indicated that increasing β values improved SNR and CNR for both non-TOF and TOF images. At a β value of 100, SNR and CNR increased across all sphere sizes (10 mm, 13 mm, 17 mm, 22 mm, 28 mm, 37 mm) when comparing non-TOF and TOF images. However, β values of 500 or higher led to decreased SNR and CNR, particularly in larger spheres (22 mm, 28 mm, 37 mm), when TOF was utilized.
Conclusion:
These findings underscore the importance of optimizing β values and employing TOF reconstruction in PET/CT scans to achieve the highest possible image quality.
Implications For Practice:
In clinical practice, practitioners should adjust β values in accordance with routine protocols, considering the size of the target region and the use of TOF reconstruction.