Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Comparison of ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based
Hansol Park1, Young-Sik Yoo2, Eunhae Shin1
1Department of Ophthalmology, Samsung Medical Center School of Medicine, Sungkyunkwan University, Seoul, Republic of Korea.
This study compared two swept-source optical coherence tomography (SS-OCT) biometers for ocular measurements and refractive prediction. While Anterion showed some measurement differences, both devices provided similar refractive prediction accuracy after cataract surgery.
Area of Science:
- Ophthalmology
- Medical Devices
- Biometry
Background:
- Comparison of two swept-source optical coherence tomography (SS-OCT) biometers: Anterion and Argos.
- Evaluation of ocular biometry agreement and postoperative refractive outcome prediction.
Purpose of the Study:
- To assess the agreement in ocular biometry measurements between Anterion and Argos SS-OCT devices.
- To evaluate the prediction accuracy of postoperative refractive outcomes using these two biometers.
Main Methods:
- Ambispective analysis of 105 eyes undergoing cataract surgery.
- Biometric measurements (axial length, central corneal thickness, lens thickness, anterior chamber depth, corneal curvature) were taken using both Anterion and Argos.
- Intraocular lens (IOL) power, mean error (ME), and mean absolute error (MAE) were calculated using standard formulas (Barrett Universal II, Haigis, Hoffer Q).
Main Results:
- Anterion yielded significantly greater axial length, central corneal thickness, and lens thickness compared to Argos.
- Anterion showed smaller keratometry (corneal curvature) measurements than Argos; no significant difference in anterior chamber depth.
- Postoperative prediction errors varied by formula, with Anterion showing a slight myopic shift tendency; however, MAE and percentage of eyes within ±0.5D were not significantly different between devices.
Conclusions:
- Despite statistically significant differences in some ocular measurements (axial length, CCT, LT, K), these were not clinically significant.
- Both Anterion and Argos demonstrated comparable refractive predictability, with no significant differences in mean absolute error.
- Anterion showed a slight tendency towards myopic outcomes, but overall accuracy within ±0.5D was similar between the two SS-OCT biometers.