Technical validation of the Zeto wireless, dry electrode EEG system.
Zoltan Nadasdy1,2,3, Adam S Fogarty4, Robert S Fisher4
1Zeto, Inc., Santa Clara, CA, United States of America.
Biomedical Physics & Engineering Express
|January 2, 2025
Summary
The Zeto WR 19 wireless EEG system demonstrates non-inferior signal quality compared to conventional EEG systems. This innovative technology offers equivalent readability and interpretability for clinical EEG applications.
Area of Science:
- Neuroscience
- Biomedical Engineering
- Medical Devices
Background:
- Clinical adoption of new EEG technology requires validation against conventional systems.
- Assessing signal quality is crucial for ensuring diagnostic accuracy and patient care.
- Zeto's WR 19 wireless dry electrode EEG system represents an innovation in EEG technology.
Purpose of the Study:
- To evaluate the signal quality of the Zeto WR 19 (zEEG) system against a conventional clinical EEG (cEEG) system.
- To determine the non-inferiority of zEEG signal quality in a clinical setting.
- To compare key signal quality metrics between the two EEG systems.
Main Methods:
- Simultaneous 30-minute EEG recordings were performed on 15 patients using both zEEG and cEEG systems.
- Signal quality was assessed using time domain statistics (Hjorth parameters), waveform correlation, spectral density, signal-to-noise ratio (SNR), and signal stability.
- Statistical comparisons were made to determine equivalence and non-inferiority.
Main Results:
- Time domain statistics showed equivalence between zEEG and cEEG, with zEEG exhibiting reduced sensitivity to electrical noise.
- Point-by-point waveform correlation was acceptable (r > 0.6, P < 0.001).
- High spectral correlation (r > 0.99, P < 0.001) and overlapping spectral density confirmed no systematic signal distortion. The zEEG system showed a 16% improvement in mean SNR (4.82 dB higher than cEEG) and maintained signal stability.
Conclusions:
- The zEEG system is non-inferior to conventional clinical EEG systems across all critical signal quality parameters.
- The Zeto WR 19 wireless dry electrode system provides signal quality equivalent to traditional cEEG recordings.
- This validates the use of the zEEG system for clinical EEG applications, ensuring readability and interpretability.