Artisse intrasaccular device for intracranial aneurysm treatment: a technical video case
Otavio F De Toledo1,2, Salvador F Gutierrez-Aguirre1,2, Montserrat Lara-Velazquez1
1Lyerly Neurosurgery, Baptist Medical Center Jacksonville, Jacksonville, Florida, USA.
Abstract:
The Artisse intrasaccular device (Medtronic) offers a novel treatment option for unruptured and ruptured wide-neck bifurcating intracranial aneurysms.1 2 The Artisse device features enhancements including a distal tip for dome protection, platinum band markers for improved visibility, and a bilayer high-density platinum core nitinol mesh basket for enhanced flexibility and visibility when compared with previous devices.2-7 Data from case series demonstrate the procedural safety and efficacy of the Artisse device.1 2 This versatile flared-shape intrasaccular device can be used alongside products such as the Rebar microcatheter 18 and Navien.2 In Europe, an adequate occlusion rate of 78% at 12 months and 79.2% at 36 months follow-up has been shown.3 8 This video documents the embolization of a wide-neck unruptured anterior communicating artery aneurysm using the Artisse device, where Sim&Size was used for proper sizing (video 1).4 9 The procedure was uneventful, achieving successful device deployment with no parent vessel compromise. neurintsurg;18/8/1838/V1F1V1Video 1Artisse intrasaccular device for intracranial aneurysm treatment-first in the USA.