"On-off" elution mechanism facilitates the rapid LC/MS/MS-based analysis of peptide antibiotics in human plasma
Abstract:
Individualized medication with peptide antibiotics, guided by therapeutic drug monitoring, is essential to treat infections caused by multidrug-resistant bacteria. Peptide antibiotics exhibit an "on-off" elution mechanism on a C18 column, leading to adsorption at the column inlet in all-aqueous conditions. Unlike small molecules, column length minimally influences their retention, with longer columns simply broadening peptide antibiotic peaks due to unnecessary post-column volume. Our theory suggests short columns can achieve comparable separation quality and enable faster analysis. Consequently, we developed a rapid LC-MS/MS method using an ultra-short (4 × 2.0 mm) column to quantify five peptide antibiotics in human plasma simultaneously. Calibration curves demonstrated strong linear regression (R2 > 0.996). The inter- and intra-accuracy at the three QC levels ranged from 86.7 % to 109.1 %, and at the LLOQ, it was between 87.6 % and 116.0 %. The precision for QCs and LLOQ was consistently below 11.7 % and 18.5 %, respectively. This method, characterized by simplicity and universality, was undoubtedly useful in clinically tailoring peptide antibiotic medication for individual patients.
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Elution Process