Effect of flow rate on plate height and resolution for antisense oligonucleotides under hydrophilic interaction
Daniel Meston1, Todd D Maloney2, Dwight R Stoll1
1Department of Chemistry, Gustavus Adolphus College, Saint Peter, MN 56082, United States.
Abstract:
Determination of quality attributes of antisense oligonucleotides (ASOs) such as purity, potency, and sequence is challenging due to their relatively large size, polyanionic nature, and large number of synthetic modifications. Chromatography technologies are evolving rapidly to meet these challenges, and one area of particularly rapid change at this time is the use of hydrophilic interaction liquid chromatography (HILIC) for oligonucleotide (ON) separations. Relatively little has been published on the factors that dictate the kinetics of these separations. This knowledge gap consequently makes it difficult to know what gains might be made during method development by changing flow rate or particle size, for example. In this work we have taken initial steps to address this gap by examining the dependence of plate height and resolution on flow rate for separations of 23-mer ASOs under HILIC conditions. Such work is complicated by the fact that the retention of these molecules decreases dramatically with increasing pressure. After adjusting mobile phase composition to hold retention factor nominally constant for each flow rate used, we find that plate height increases strongly with increasing flow rate such that the plate height increases about ten-fold over the range of flow rate of 0.1 to 4.0 mL/min. when using a 4.6 mm i.d. column. However, the minimum reduced plate height observed at the lowest flow rate is quite impressive at around 2. Finally, we find that this dependence of plate height on flow rate translates, as expected, to an improvement in resolution as flow rate is decreased, both in conventional one-dimensional separations, and in the second dimension of a two-dimensional separation. We expect to use this work as a foundation to build on as we deepen our understanding of the kinetics of ON separations.
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...