Prototype analysis of a low-power, small-scale wearable medical device
Pablo Dutra da Silva1, Pedro Bertemes Filho1
1Electrical Engineering Department, State University of Santa Catarina, Santa Catarina, Brazil.
Abstract:
Wearable and portable devices are gaining significant popularity across consumer electronics as well as in medical and industrial fields. To ensure that these devices are both comfortable and appealing to users, they need to have low battery consumption and be compact in both size and weight. The EGluco project is focused on developing a wearable device for non-invasive blood glucose monitoring. This multi-sensor device incorporates electrical bioimpedance spectroscopy as one of its measurement techniques. One of the earlier versions of the device was deemed unsuitable as a wearable due to its large size and high power consumption. To make the device more suitable for wearability, the previous hardware was assessed, and a new design was proposed that simplified the system's power supply and reduced the operating voltage. This article presents two of these designs: an improved Howland current source with a supply voltage of 3.3 V, an output current of 250 μA, and the ability to conduct bioimpedance analysis up to 1 MHz using pulsed DIBS (Discrete Interval Binary Sequence) signals, and an instrumentation amplifier with the same supply voltage as the current source, a voltage gain of four, and a slew rate of 150 V/μs. By simplifying the power supply and implementing other changes, the device's size was reduced to a single 5 × 5 cm circuit board, compared to the previous configuration of four separate boards connected by cables.