Quantifying X-Ray Fluorescence Data Using MAPS
Published on: February 17, 2018
Preliminary studies for the standardization of a pXRF analyzer via ICP-OES for the accurate quantification of Pb in
Caryn S Seney1, Adam M Kiefer1, Samantha T Brown1
1Department of Chemistry, Mercer University, Macon, GA 31201, USA. kiefer_am@mercer.edu.
Abstract:
Determining lead (Pb) concentrations in new paints using spectroscopic methods such as Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES) requires technical expertise, consumables, equipment for method preparation, and instrumentation that can be cost prohibitive and difficult to maintain in low and middle-income countries (LMICs). Although portable X-ray Fluorescence (pXRF) analyzers are less expensive and simple to operate, their inaccuracy has limited their use to screening for the analysis of Pb in new, dried paint. To determine the limits of pXRF analyzers, new paint samples were purchased, dried, homogenized, and analyzed via pXRF and ICP-OES. Pb concentrations determined via pXRF were ∼27% lower than those determined by ICP-OES. Interestingly, the concentrations determined by both methods exhibited a strong linear correlation. The resulting equations were used to calibrate an individual pXRF analyzer, improving the accuracy of the pXRF result to within ±4.4% of verified concentrations of two certified reference materials. Additionally, eleven new paint samples from Benin and Türkiye were analyzed via pXRF, and their calculated concentrations were similar within ±10% of concentrations as verified by ICP-OES. This method requires no acid digestion or dilutions and provides an alternative method for the determination of Pb in new paint in LMICs.
More Related Videos
14:53In Situ Detection and Single Cell Quantification of Metal Oxide Nanoparticles Using Nuclear Microprobe Analysis
Published on: February 3, 2018
08:12Characterization, Quantification and Compound-specific Isotopic Analysis of Pyrogenic Carbon Using Benzene Polycarboxylic Acids BPCA
Published on: May 16, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Emission Spectroscopy: Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences