Impact of leadless pacemaker implantation site on cardiac synchronization and tricuspid regurgitation
Xiao-Fei Gao1, Hong Zhu2, Jia-Sheng Zhang1
1Department of Cardiology, Hangzhou First People's Hospital, #261 Huansha Road, Hangzhou, 310000, Zhejiang Province, China.
Background:
To investigate the optimization of leadless pacemaker placement and to assess its impact on heart synchronization and tricuspid regurgitation.
Results:
A clinical trial was conducted involving 53 patients who underwent leadless pacemaker implantation at the Second Affiliated Hospital of Zhejiang University School of Medicine and Hangzhou First People's Hospital between March 2022 and February 2023. Implantation site localization was determined using the 18-segment method under RAO 30° imaging. Intraoperative and 1-month post-operative echocardiography was performed to assess cardiac electromechanical synchronization and tricuspid regurgitation; parameters of interest included interventricular mechanical delay (IVMD), pre-ejection period of the aorta (L-PEI), and septal-to-posterior wall motion delay (SPWMD). Pacing thresholds, sensing, and impedance exhibited no significant differences between the 8/9 zone and other sites (P > 0.05). In contrast, the 8/9 zone group manifested a significant reduction in L-PEI (128.24 ± 12.27 vs. 146.50 ± 18.17 ms, P < 0.001), IVMD (17.92 ± 8.47 vs. 28.56 ± 15.16 ms, P < 0.001), and SPWMD (72.84 ± 19.57 vs. 156.56 ± 81.54 ms, P < 0.001), compared to the non-8/9 group. Post-pacing QRS duration showed no significant difference between the two groups (139.21 ± 11.36 vs. 143.83 ± 16.35 ms P = 0.310). Notably, for patients with atrial fibrillation, the 8/9 zone placement significantly reduced tricuspid regurgitation. During the 1-month follow-up, neither group reported major complications such as bleeding, cardiac tamponade, pacemaker detachment, or malignant arrhythmias.
Conclusion:
Implantation of the leadless pacemaker in the right ventricular 8/9 zone provides superior electromechanical synchronization compared to other sites.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization