Objective:
The aim of this study is to develop a set of management tools for early warning of steam sterilization failure, including a failure risk checklist and a calculation model for assessing early sterilization failure risk, and to verify the early warning effectiveness of the management tools through check experiment.
Methods:
This study included two stages. The first stage involved the construction of a failure risk checklist and the development of a computational formula for early sterilization failure risk by expert consultation. In the second stage, the early warning effectiveness of the sterilization failure management tools was verified through a comparative verification method based on the consultation results of two experts.
Results:
The evaluation system comprised a total of 10 early warning indicators divided into 3 categories: "Staff", "Steam sterilizer" and "Medical devices". The Cronbach's α coefficient for internal consistency reliability was 0.837, the total content validity index (S-CVI) was 0.884, and the item-content validity (I-CVI) ranged from 0.762 to 1. In management practices, the sterilization qualification rate of the experimental group was 100%, compared to 99.78% in the control group. There were significant differences between the two groups in terms of accidental downtime (hours) and the normal operation rate of the equipment caused by equipment failure.
Conclusion:
This study constructed an early warning model for the vacuum steam sterilization process failures. This model facilitates more accurate identification of sterilization failure risks, which helps medical institutions in enhancing sterilization quality control and early interventions.
More Related Videos
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...