Simultaneous reticulocyte and platelet counting on a clinical flow cytometer
American Journal of Hematology
|April 1, 1985
Summary
Automated laser flow cytometry using acridine orange provides efficient, rapid, and reliable reticulocyte and platelet counts. This method shows high correlation with standard techniques, offering consistent results, especially for marrow transplant patients.
Area of Science:
- Hematology
- Clinical Immunology
- Biomedical Engineering
Background:
- Laser flow cytometry is a versatile tool for clinical immunofluorescence and hematologic analyses.
- Automated enumeration of reticulocytes and platelets is a key application in hematology.
Purpose of the Study:
- To clinically evaluate a fluorescence laser flow cytometer method for automated reticulocyte and platelet enumeration.
- To assess the accuracy and reliability of this method using the fluorochrome acridine orange.
Main Methods:
- Analysis of 132 clinical specimens for platelet counts and 101 for reticulocyte counts using a fluorescence laser flow cytometer.
- Simultaneous enumeration by conventional techniques for comparative analysis.
- Utilized the fluorochrome acridine orange for cell staining.
Main Results:
- High correlation (R = .989 for reticulocytes, R = .995 for platelets) with standard methods in non-bone marrow transplant patients.
- Slightly lower correlation (R = .923) for reticulocyte enumeration in marrow transplant patients.
- Automated method demonstrated more consistent serial reticulocyte count results over time compared to manual methods in transplant patients.
Conclusions:
- Fluorescence laser flow cytometry with acridine orange is an efficient, rapid, and reliable method for platelet and reticulocyte counting.
- The method is well-suited for clinical hematologic applications, including in cancer centers.
- Offers improved consistency for serial monitoring in complex patient populations like marrow transplant recipients.