Reliability and Sensitivity of Enode/Vmaxpro Sensor for Muscle Power Assessment
Milan Marković1, Lazar Toskić1,2, Veroljub Stanković1,3
1Faculty of Sport and Physical Education, University of Priština-Kosovska Mitrovica, 38218 Leposavic, Serbia.
Life (Basel, Switzerland)
|January 8, 2025
Summary
The Enode/Vmaxpro sensor reliably measures muscle power, showing high accuracy in velocity and power metrics for athletes. This device is suitable for sports diagnostics and biomechanical analysis.
Area of Science:
- Biomechanics
- Sports Science
- Human Movement Analysis
Background:
- Performance monitoring is crucial for training, irrespective of physical activity level.
- The Enode/Vmaxpro sensor is a tool used for assessing various performance parameters.
- Evaluating the sensor's reliability and sensitivity is essential for its practical application.
Purpose of the Study:
- To assess the reliability and sensitivity of the Enode/Vmaxpro sensor for performance parameter measurements.
- To determine the metric characteristics of average velocity, peak velocity, average power, peak power, movement length, and movement duration.
- To validate the sensor's utility in biomechanics and sports diagnostics.
Main Methods:
- Twenty-seven participants (15 men, 12 women) underwent body composition analysis.
- Participants performed leg extension/leg curl exercises with each leg under a 30% body mass load.
- Descriptive statistics, reliability analyses (ICC), and difference analyses were used to evaluate repeatability and sensitivity.
Main Results:
- High reliability was found for the Enode/Vmaxpro sensor across all participants (ICC: 0.937-0.991), men (ICC: 0.899-0.984), and women (ICC: 0.908-0.990).
- Significant differences in performance were observed based on sex, activity level, and movement type (p=0.000).
- No significant differences were found between right and left leg performance.
Conclusions:
- The Enode/Vmaxpro sensor exhibits adequate reliability for muscle power testing.
- The sensor demonstrates sufficient sensitivity to detect differences in performance.
- The findings support the use of the Enode/Vmaxpro sensor in biomechanics and sports diagnostics.