Characterization of an Inorganic Powder-Based Scintillation Detector Under a UHDR Electron Beam
Daline Tho1, Sam Beddar1,2
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA.
Abstract:
(1) Background: Ultra-high dose rate (UHDR) radiation therapy needs a reliable dosimetry solution and scintillation detectors are promising candidates. In this study, we characterized an inorganic powder-based scintillation detector under a 9 MeV UHDR electron beam. (2) Methods: A mixture of ZnS:Ag powder and optic glue was coupled to an 8 m Eska GH-4001-P polymethyl methacrylate (PMMA) optical fiber. We evaluated the dependence of the detector on dose per pulse (DPP), pulse repetition frequency (PRF), and pulse width (PW). Additionally, we determined the stability and the reproducibility of the detector. (3) Results: The signal ratio between the PMMA clear optical fiber and the ZnS:Ag scintillator was around 210. ZnS:Ag produced a signal yield 54 times greater than that of a BCF-12 plastic scintillator. Signal variation with PRF changes was under 0.5%. The signal was linear to the integrated dose up to the maximum deliverable dose, 180 Gy. The variation in signal was linear to the change in both PW and DPP. Regarding stability, the standard deviation of 10 consecutive irradiations was 0.83%. For the reproducibility, all daily measurements varied within ±1.5%. (4) Conclusions: These findings show that the ZnS:Ag detector can be used for accurate dosimetry with UHDR beams.
More Related Videos
11:26Integrating a Triplet-triplet Annihilation Up-conversion System to Enhance Dye-sensitized Solar Cell Response to Sub-bandgap Light
Published on: September 12, 2014
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Gas Chromatography: Types of Detectors-II
UV–Vis Spectrometers
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...