Related Experiment Video
Updated: May 3, 2026

Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers
Petr Neuzil1, Derek V Exner2, Reinoud E Knops3
1Na Homolce Hospital, Prague, Czech Republic.
Background:
The growing use of leadless pacemaker (LP) technology requires safe and effective solutions for retrieving and removing these devices over the long term.
Objectives:
This study sought to evaluate retrieval and removal of an active helix-fixation LP studied in worldwide regulatory clinical trials.
Methods:
Subjects enrolled in the LEADLESS II phase 1 investigational device exemption, LEADLESS Observational, or LEADLESS Japan trials with an attempted LP retrieval at least 6 weeks postimplantation were included. The retrieval success rate was evaluated for all attempted retrievals. Site-reported complications associated with LP retrievals were quantified, with the severity and relationship adjudicated by an independent clinical events committee.
Results:
There were 1,423 successful LP implants and 234 chronic LP retrieval attempts in 233 subjects. The overall retrieval success rate was 87.6% (205 of 234 attempts). The most common reason for failed retrieval (86.2%, 25 of 29 attempts) was inability to access the LP docking button. The mean time from implantation to attempted retrieval was 3.2 years (range 0.2-9.0 years) and did not significantly influence the retrieval success (P = 0.71). The clinical events committee adjudicated 11 complications in 9 subjects as being retrieval related (3.9%, 9 of 233 subjects).
Conclusions:
The safety and efficacy of chronic LP retrieval of an active helix-fixation device was demonstrated in an international clinical trial setting, with implant durations up to 9 years.
Related Concept Videos
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care