Measuring Cardiac Autonomic Nervous System ANS Activity in Children
Published on: April 29, 2013
Repeatability of RRate measurements in children during triage in two Ugandan hospitals
Ahmad Asdo1,2, Alishah Mawji1,2, Isaac Omara3
1Department of Anesthesiology, Pharmacology & Therapeutics, University of British Columbia, Vancouver, British Columbia, Canada.
Insights
The RRate app demonstrated high repeatability and efficiency in measuring children's respiratory rates in Ugandan hospitals. This supports its use in low- and middle-income countries for accurate pneumonia diagnosis.
Area of Science:
- Pediatrics
- Global Health
- Medical Technology
Background:
- Pneumonia is a leading cause of child mortality globally, particularly in low- and middle-income countries (LMICs).
- Accurate respiratory rate assessment is crucial for diagnosing pneumonia but can be unreliable with less-trained healthcare workers.
- The RRate app aims to improve the accuracy and reliability of respiratory rate measurements.
Purpose of the Study:
- To investigate the repeatability of the RRate app for measuring respiratory rates in children.
- To assess the efficiency and clinical utility of the RRate app in Ugandan district hospitals.
Main Methods:
- A secondary analysis of 3,679 children aged 0-5 years using sequential respiratory rate measurements from 14 nurses.
- Measurements were taken using the RRate app, with analysis focused on agreement between paired observations.
- Intraclass Correlation Coefficient (ICC) was used to assess measurement repeatability, with data clustered around WHO fast-breathing thresholds.
Main Results:
- Respiratory rate measurements were efficient, taking less than 15 seconds for 98.9% of readings.
- Breathing classification changed between sequential measurements in only 12.6% of children.
- High repeatability was observed, with an overall ICC of 0.95 (0.94-0.95) and 12 of 14 users achieving ICC ≥ 90%.
Conclusions:
- The RRate app provides efficient and highly repeatable respiratory rate measurements.
- Its simplicity and reliability support its adoption in LMIC healthcare settings for improved pneumonia diagnosis.
- Widespread clinical adoption of the RRate app is endorsed for enhancing pediatric respiratory assessments.
Abstract:
Pneumonia is the leading cause of death in children globally. In low- and middle-income countries (LMICs) pneumonia diagnosis relies on accurate assessment of respiratory rate, which can be unreliable when completed by nurses with less-advanced training. To inform more accurate measurements, we investigate the repeatability of the RRate app used by nurses in Ugandan district hospitals. This secondary analysis included 3,679 children aged 0-5 years. The dataset had two sequential measurements of respiratory rate collected by 14 nurses using the RRate app. We measured agreement between respiratory rate observations while indicating observations' clustering around WHO fast-breathing thresholds. WHO thresholds are 60 breaths per minute (bpm) for under two months (Age-1), 50 bpm for two to 12 months (Age-2), and 40 bpm for 12.1 to 60 months (Age-3). We assessed the repeatability of the paired measurements per user through the Intraclass Correlation Coefficient (ICC) and calculated an overall ICC value. The respiratory rate measurement took less than 15 seconds for 7,277 (98.9%) of the measurements. Despite respiratory rates clustering around WHO thresholds, breathing classification based on the thresholds (Fast vs normal) was altered between sequential measurements in only 12.6% of children. The mean (SD) respiratory rate by age group was 60 (13.1) bpm for Age-1, 49 (11.9) bpm for Age-2, and 38 (10.1) for Age-3, and the bias (Limits of Agreements) were 0.3 (-10.8-11.3) bpm, 0.4 (-8.5-9.3) bpm, and 0.1 (-6.8, 7.0) bpm for Age-1, Age-2, and Age-3 respectively. The repeatability of the paired respiratory rate measurements was high, with an ICC ≥ 90% for 12 of 14 users and an overall ICC value (95% CI) of 0.95 (0.94-0.95). The RRate measurements were efficient and repeatable. The simplicity, repeatability, and efficiency support its usage in LMICs healthcare facilities, and endorses a more widespread clinical adoption.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of blood pressure in brachial artery(two-step method)