New gantry angle-dependent beam control optimization with Elekta linear accelerator for VMAT delivery
Adriaan Abraham van Appeldoorn1, Johannes Gerardus Maria Kok1, Jochem Willem Heiko Wolthaus1
1Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands.
Introduction:
This paper describes a method to improve gantry-dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions. These interruptions are known to contribute to unwanted intrafraction motion and longer treatment times.
Methods:
Compared to the default method of the manufacturer, this new method takes both clockwise and counterclockwise rotation to compensate for magnetic hysteresis as well as previous configuration and noise filtering into account. The improved method to determine the lookup table uses service graphing information from the linac without the need for additional symmetry information. The clinical configuration of the flattened beam energies remains untouched during the data record.
Results:
This method results in a configuration where the gantry-dependent steering is optimized over the full arc with optimal balance in the hysteresis and minimizing the effect of errors in the steering values. This method is a less error-prone process compared to the methodology described in previous research, still achieving a reduction of interruption of about 60 percent compared to the Elekta method.
Conclusion:
This study shows a simplified way to optimize linac stability with improved LUT. The optimized LUT results in a lower number of interruptions, preventing downtime, and a lower risk of intrafraction motion due to longer treatment time.
Related Concept Videos
Angle Closure Glaucoma: Treatment
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal centroidal axes. The...
Shear on the Horizontal Face of a Beam Element
Beams with Symmetric Loadings
The M/EI...
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...