Comparison between the SGRT and the conventional setup method for patients undergoing VMAT for pelvic malignancies
V Sotiropoulou1, F Tsironi1, M Tolia2
1Department of Medical Physics, Medical School, University of Crete, Heraklion, Greece.
Summary
Surface Guided Radiation Treatment (SGRT) offers more accurate patient positioning for pelvic cancer radiotherapy compared to traditional laser and tattoo methods. While SGRT increases setup time, it significantly reduces the need for position corrections during treatment.
Area of Science:
- Radiation Oncology
- Medical Physics
- Cancer Treatment
Background:
- Accurate patient positioning is critical for effective radiotherapy, especially for pelvic malignancies.
- Conventional methods using lasers and tattoos have limitations in precision and consistency.
- Surface Guided Radiation Treatment (SGRT) utilizes visual light cameras for real-time surface monitoring.
Purpose of the Study:
- To compare the accuracy and time efficiency of SGRT versus the conventional laser/tattoo method for patient positioning.
- To evaluate SGRT's effectiveness in patients with pelvic malignancies undergoing radiotherapy.
Main Methods:
- A study involving 34 patients (24 male prostate, 10 female gynecological) undergoing radiotherapy.
- Patients were divided into two groups: Group A (tattoo/laser) and Group B (SGRT), with 17 patients each.
- Position verification using cone beam computed tomography (CBCT) before each session; analysis of positional corrections and setup time.
Main Results:
- SGRT (Group B) demonstrated significantly smaller positional corrections across all directions (vertical, lateral, longitudinal) compared to the conventional method (Group A).
- Improvements in corrections were 16.6% (vertical), 36.7% (lateral), and 33.3% (longitudinal), with a 29.2% improvement in total corrections.
- SGRT showed greater consistency (IQR 0.25 cm vs 0.38 cm) but required a significantly longer median setup time (2.0 min vs 1.6 min).
Conclusions:
- Surface Guided Radiation Treatment (SGRT) enhances positioning accuracy for Volumetric Modulated Arc Therapy (VMAT) in pelvic cancer patients.
- Despite a longer setup time, SGRT provides superior precision compared to the conventional laser and tattoo-based positioning techniques.