Reproducibility and repeatability of 18F-(2S, 4R)-4-fluoroglutamine PET imaging in preclinical oncology models.
Gregory D Ayers1,2, Allison S Cohen3,4,5, Seong-Woo Bae5
1Department of Biostatistics, Vanderbilt University Medical Center, Nashville, TN, United States of America.
This study demonstrates high repeatability and reproducibility for the PET tracer 18F-Gln in cancer research. Appropriate statistical methods confirm the reliability of 18F-Gln measurements for in vivo studies.
Area of Science:
- Nuclear medicine
- Molecular imaging
- Oncology research
Background:
- Positron emission tomography (PET) repeatability and reproducibility (R&R) are crucial for its clinical potential.
- While R&R for 18F-FDG PET is established, data for other tracers like 18F-(2S, 4R)-4-fluoro-glutamine (18F-Gln) are limited.
- This study addresses the need for R&R assessment of 18F-Gln, a tracer for glutamine uptake and metabolism.
Purpose of the Study:
- To evaluate the repeatability and reproducibility of 18F-Gln PET imaging.
- To assess the agreement and R&R using statistical methods specifically designed for this purpose.
- To demonstrate the reliability of 18F-Gln for in vivo research applications.
Main Methods:
- Twenty colorectal cancer xenografts in mice were imaged using 18F-Gln PET.
- Three analysts performed tumor uptake analysis on the same images.
- Repeatability was assessed after 6 months, with scans re-ordered.
- Statistical analyses included Bland and Altman (B&A), Gauge Reproducibility and Repeatability (Gauge R&R), and Lin's Concordance Correlation Coefficient.
- An equivalency test was conducted to confirm agreement.
Main Results:
- Gauge R&R indicated that reproducibility and repeatability accounted for 31% and 69% of total measurement error, respectively.
- Bland and Altman analysis showed good repeatability coefficients and agreement limits among analysts.
- Equivalency tests confirmed that repeated measures fell within predefined equivalence limits (± 0.5%ID/g).
Conclusions:
- The study demonstrates high individual analyst and laboratory-level reproducibility and repeatability for 18F-Gln PET.
- Appropriate statistical methods for R&R assessment are critical and should be widely adopted.
- These findings support the use of 18F-Gln for reliable in vivo research.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
06:51Utilizing 18F-FDG PET/CT Imaging and Quantitative Histology to Measure Dynamic Changes in the Glucose Metabolism in Mouse Models of Lung Cancer
Published on: July 21, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...