An evaluation of a leukocyte esterase/nitrite test strip and a bioluminescence assay for detection of bacteriuria
Related Experiment Video
Updated: Aug 18, 2026

10:05
Colorimetric Detection of Bacteria Using Litmus Test
Published on: September 17, 2016
Abstract:
A leukocyte esterase/nitrite (Chemstrip LNTM) and a bioluminescence assay (LumacTM) were evaluated for detection of bacteriuria in 1,000 urine specimens. Both devices provided high predictive negative values (95.4-97.8%); however, false-negative values at levels of bacteriuria greater than or equal to 10(4) CFU/ml were 22.6% and 12.3%, respectively, for the leukocyte esterase/nitrate and the bioluminescence assay tests. The corresponding false-negative values at levels of bacteriuria of greater than or equal to 10(5) CFU/ml were 16.5% and 3.9%, respectively.