Assessment of the Performance of the Dose Calibrator Used in Radioactivity Measurement
Dinh Xuan Hoang1, Truong Dinh Vu1, Do Van Dan1
1Center for Radiation Protection, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.
Aims:
This study aimed to evaluate the principal technical characteristics of a well-type gas-filled ionization chamber dose calibrator used in measuring radiopharmaceutical activity, namely accuracy, repeatability, and linearity. Furthermore, this work also explored the correlation between the device's response and the position and volume of the radiopharmaceutical I-131.
Materials And Methods:
Experimental measurements were conducted on the ATOMLAB 500 dose calibrator using NIST traceable Cs-137 source to determine the accuracy and repeatability. For the linearity test, the Tc-99m solution produced at the Dalat Nuclear Research Institute, Vietnam was utilized to examine the performance of the device over a wide measurement range. Effects resulting from different volumes and measuring positions were also determined by experiments and Monte Carlo simulations.
Results:
Based on acceptance test results, it is revealed that the dose calibrator's characteristics comply with international standards, where the deviations of accuracy, repeatability, and linearity are all lower than 1.0%. The response of the dose calibrator to different measuring positions and volumes was well controlled. To rectify the discrepancy in the response to the changes in solution volume, correction functions were proposed.
Conclusions:
It is concluded that the dose calibrator is suitable for radioactivity quantification and adheres to standards recommended by international organizations for nuclear medicine activities. The outcome of this work will serve as the foundation for establishing a standard for I-131 radiopharmaceuticals' radioactivity in Vietnam.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Biological Effects of Radiation
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Constant Pressure Calorimetry