Comparative Evaluation of Consumer Wearable Devices for Atrial Fibrillation Detection: Validation Study
Femke Wouters1,2, Henri Gruwez1,2,3,4, Christophe Smeets1,2
1Limburg Clinical Research Center/Mobile Health Unit, Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium.
Background:
Consumer-oriented wearable devices (CWDs) such as smartphones and smartwatches have gained prominence for their ability to detect atrial fibrillation (AF) through proprietary algorithms using electrocardiography or photoplethysmography (PPG)-based digital recordings. Despite numerous individual validation studies, a direct comparison of interdevice performance is lacking.
Objective:
This study aimed to evaluate and compare the ability of CWDs to distinguish between sinus rhythm and AF.
Methods:
Patients exhibiting sinus rhythm or AF were enrolled through a cardiology outpatient clinic. The participants were instructed to perform heart rhythm measurements using a handheld 6-lead electrocardiogram (ECG) device (KardiaMobile 6L), a smartwatch-derived single-lead ECG (Apple Watch), and two PPG-based smartphone apps (FibriCheck and Preventicus) in a random sequence, with simultaneous 12-lead reference ECG as the gold standard.
Results:
A total of 122 participants were included in the study: median age 69 (IQR 61-77) years, 63.9% (n=78) men, 25% (n=30) with AF, 9.8% (n=12) without prior smartphone experience, and 73% (n=89) without experience in using a smartwatch. The sensitivity to detect AF was 100% for all devices. The specificity to detect sinus rhythm was 96.4% (95% CI 89.5%-98.8%) for KardiaMobile 6L, 97.8% (95% CI 91.6%-99.5%) for Apple Watch, 98.9% (95% CI 92.5%-99.8%) for FibriCheck, and 97.8% (95% CI 91.5%-99.4%) for Preventicus (P=.50). Insufficient quality measurements were observed in 10.7% (95% CI 6.3%-17.5%) of cases for both KardiaMobile 6L and Apple Watch, 7.4% (95% CI 3.9%-13.6%) for FibriCheck, and 14.8% (95% CI 9.5%-22.2%) for Preventicus (P=.21). Participants preferred Apple Watch over the other devices to monitor their heart rhythm.
Conclusions:
In this study population, the discrimination between sinus rhythm and AF using CWDs based on ECG or PPG was highly accurate, with no significant variations in performance across the examined devices.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...