Nyquist ghost elimination for diffusion MRI by dual-polarity readout at low b-values
Oscar Jalnefjord1,2, Nicolas Geades3, Guillaume Gilbert4
1Department of Medical Radiation Sciences, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.
Abstract:
Dual-polarity readout is a simple and robust way to mitigate Nyquist ghosting in diffusion-weighted echo-planar imaging but imposes doubled scan time. We here propose how dual-polarity readout can be implemented with little or no increase in scan time by exploiting an observed b-value dependence and signal averaging. The b-value dependence was confirmed in healthy volunteers with distinct ghosting at low b-values but of negligible magnitude atb= 1000 s/mm2. The usefulness of the suggested strategy was exemplified with a scan using tensor-valued diffusion encoding for estimation of parameter maps of mean diffusivity, and anisotropic and isotropic mean kurtosis, showing that ghosting propagated into all three parameter maps unless dual-polarity readout was applied. Results thus imply that extending the use of dual-polarity readout to low non-zero b-values provides effective ghost elimination and can be used without increased scan time for any diffusion MRI scan containing signal averaging at low b-values.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...