Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Effect of Patient Positioning on CT Number Accuracy: A Phantom Study Comparing Energy Integrating and Deep Silicon
Aria M Salyapongse1, Sean D Rose2, Perry J Pickhardt
1Department of Radiology, University of WisconsinâMadison, Madison, WI.
Objective:
Patient positioning during clinical practice can be challenging, and mispositioning leads to a change in CT number. CT number fluctuation was assessed in single-energy (SE) EID, dual-energy (DE) EID, and deep silicon photon-counting detector (PCD) CT over water-equivalent diameter (WED) with different mispositions.
Methods:
A phantom containing five clinically relevant inserts (Mercury Phantom, Gammex) was scanned on a clinical EID CT and a deep silicon PCD CT prototype at vertical positions of 0, 4, 8, and 12 cm. EID scans used 120 kV and rapid kV-switching DE techniques. CT number was calculated for air, water, polystyrene, iodine 10 mg/mL, and bone. Ideal CT numbers were calculated using the NIST XCOM database toolkit. Comparison of measured to ideal CT number utilized relative root mean square error (RMSE). Trends in CT number versus WED were compared using linear regression and statistical comparisons to test for differences in slope.
Results:
No statistical difference of CT number with mispositioning was seen between acquisition modes. CT number fluctuation was larger due to WED than mispositioning for all material inserts. Water, iodine, and bone, for deep silicon PCD CT had statistically significant ( P < 0.05) smaller slopes compared to EIDof CT number over WED for all tested mispositions. The accuracy of deep silicon PCD CT was higher than either SE or DE EID CT for all materials at all mispositions except for polystyrene.
Conclusions:
WED (ie, patient size) contributes to CT number fluctuation more than mispositioning. The change in CT number was significantly smaller, and CT number accuracy was higher for deep silicon PCD CT in this phantom study.
More Related Videos
06:42Plasmonic Photothermal Cancer Therapy: Nanoparticle-embedded Tumor-tissue-mimicking Phantoms for Visualizing Photothermal Temperature Distribution
Published on: May 9, 2025
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...