A rapid and highly sensitive fluorescence immunochromatographic test strip for pepsin detection in human hypopharyngeal saliva
Related Experiment Video
Updated: Jun 3, 2025

The α-test: Rapid Cell-free CD4 Enumeration Using Whole Saliva
Published on: May 16, 2012
Yaqi Liu1, Lihong Chen1, Jingtian Huang1
1Otolaryngology, Head and Neck Surgery Department of Fujian Provincial Hospital, Provincial Clinical Medical College of Fujian Medical University, Fuzhou University Affiliated Provincial Hospital, Fuzhou, Fujian, 35001, China. 9201551173@fjmu.edu.cn.
Abstract:
Pepsin serves as a potential biomarker for laryngopharyngeal reflux disease (LPRD). The reported methods for detecting pepsin still have limitations, such as long reaction times and low sensitivity. Therefore, it is crucial to develop a rapid, simple and sensitive detection method. In this study, a fluorescence immunochromatographic test strip was designed for point-of-care diagnostics of LPRD. To obtain optimal fluorescence immunochromatographic conditions, key experimental factors, including microsphere particle size, binding pH, and resuspension dilution ratio, were systematically evaluated. The optimal fluorescent microsphere size was determined to be 300 nm, the most suitable reaction pH was 8.0, and the ideal resuspension solution to storage solution ratio was 1 : 4. Under these optimal conditions, the sensor had a linear range of 2.5-100.0 ng mL-1 and a detection limit of 1.9 ng mL-1. The relative standard deviation was below 15% for both intra- and inter-batch reproducibility. Under accelerated damage conditions at 37 °C, the test value of the test strip could be kept basically unchanged for 14 days. There was no cross-reactivity of the test strip with pepsinogen I, pepsinogen II and other interferents. In the actual sample detection, the recovery rate was between 85% and 115%. Furthermore, the proposed method was demonstrated to be green and to exhibit strong applicability based on the evaluation using the AGREEprep and the BAGI. In addition, the method has the advantages of high sensitivity, accuracy, greenness and applicability, which can meet the needs of rapid quantitative on-site detection.
More Related Videos
08:50Multiplexed Fluorescent Microarray for Human Salivary Protein Analysis Using Polymer Microspheres and Fiber-optic Bundles
Published on: October 10, 2013
08:37Sampling Human Indigenous Saliva Peptidome Using a Lollipop-Like Ultrafiltration Probe: Simplify and Enhance Peptide Detection for Clinical Mass Spectrometry
Published on: August 7, 2012