Development and Internal Validation of an AI-Enabled Cuff-less, Non-invasive Continuous Blood Pressure Monitor Across
Francisco Lopez-Jimenez1, Abhishek Deshmukh1, John Bisognano2
1Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN, 55905, USA.
Background:
Non-invasive, continuous blood pressure monitoring technologies require additional validation beyond standard cuff-based methods. This study evaluates a non-invasive, multiparametric wearable cuffless blood pressure (BP) diagnostic monitor across all hypertension classes with diverse subjects.
Methods:
A prospective, multicenter study assessed Nanowear's SimpleSense-BP performance, including induced and natural BP changes, significant BP variations (Systolic BP (SBP) ≥ ± 15 mm Hg and Diastolic BP (DBP) ≥ ± 10 mm Hg), and reference input value validity over 4 weeks.
Results:
303 subjects (18-83 yrs; 50.16% Female) participated in algorithmic development and validation (Normal - 35%, Prehypertensive - 24%, Stage 1 - 24%, Stage 2 - 17%). 54 subjects were tested for induced change performance, 149 exhibited significant changes, and 91 validated reference value duration.
Conclusions:
The study clinically validated a continuous, AI-based BP diagnostic monitor using non-invasive wearable data. Further testing on diverse populations and external validation are recommended. The protocol was inspired by ISO 81060-2 and IEEE 1708:2019 standards.
More Related Videos
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure