Validation Status of Electronic Sphygmomanometers in China: A National Survey
Xianghui Zhang1,2, Zhimin Xu3, Yingchun Yao4
1Department of Preventive Medicine, Shihezi University, Xinjiang, China (X.Z.).
Background:
Although the information on the validation status of electronic sphygmomanometer (ES) devices in use in health care institutions and households is much more clinically relevant than that of ES models available on the market, it remains insufficient.
Methods:
A national survey was conducted across all administrative regions of mainland China to assess the validation status of ESs. Fifty-eight cities were selected with stratification by municipality, provincial capital, and other cities, and health care institutions and households in each city were chosen by convenience to identify ES devices in use according to the study protocol. Information on devices' model, type, was collected. The validation status of each device was ascertained through searching international online registries of its models.
Results:
A total of 3230 ES devices (1924 from health care institutions and 1306 from households), encompassing 498 ES models, were actually selected. The weighted proportion of accuracy-validated devices was significantly higher in health care institutions (33.9% [95% CI, 31.8-36.0%]) compared with households (23.0% [95% CI, 20.8-25.4%]; P<0.001), and both exceeded the proportions for models (17.7% [95% CI, 13.7-22.6%] and 12.5% [95% CI, 9.5-16.3%], respectively). Regional economic development, device type, and years in use were associated with the proportion of accuracy-validated devices. Notably, 8.8% of devices used in secondary hospitals were wrist type.
Conclusions:
The accuracy validation status of ESs in China in the number of devices in use shows a more favorable situation than that in the number of models but remains low and concerning. Policies are urgently needed to promote the widespread adoption of accuracy-validated models among all users.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure