Assessing Mean Glandular Dose in Mammography in Jordan According to American College of Radiology (ACR) Standards
Rasha Al Hrout1,2, Mohammad Ayasrah3, Nurul Zahirah Noor Azman2
1Medical Radiologic Technologies, College of Allied Medical Sciences, Zarqa University, Zarqa, 13132, Jordan.
Introduction:
Breast cancer is a significant worldwide health issue, particularly in Jordan, where early detection via mammography is essential for effective disease management. Despite the little radiation risk associated with mammography, it is crucial to monitor radiation exposure to guarantee patient safety. This study intends to assess skin entrance exposure and compute the Mean Glandular Dose (MGD) in mammography units to determine adherence to established criteria and pinpoint areas for enhancement.
Methods:
To assess MGD, the study utilized the American College of Radiologists (ACR) phantom alongside a RaySafe X2 MAM dosimeter. Measurements of entrance kerma and half-value layer (HVL) were taken across 25 mammography units in Jordan. The MGD was calculated according to the ACR's 2018 protocol, which provides a standardized approach to ensure accurate and comparable dose estimations. These measurements were then analyzed against the ACR's threshold of 3 mGy to assess compliance.
Results:
The study found that the average MGD across all units was 2.3 mGy, with individual values ranging from 0.95 to 4.10 mGy. Although 67% of the units maintained MGD values within the ACR threshold, 33% exceeded the recommended limit of 3 mGy. Higher MGD values were particularly common in non-accredited facilities, where the average MGD reached 2.7 mGy, compared to 1.6 mGy in accredited units, suggesting gaps in quality control and adherence to best practices in non-accredited centers.
Conclusion:
This study emphasizes the critical role of accreditation and adherence to quality standards in maintaining safe and effective mammography practices. While most mammography units in Jordan meet the ACR's recommended MGD limits, the elevated dose levels in some non-accredited facilities highlight the need for more rigorous implementation of accreditation standards. Improving compliance with established guidelines will enhance breast cancer screening effectiveness, ultimately supporting better early detection and outcomes for breast cancer in Jordan.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT