Wilson S Tsai1, Erin Haywood1, Xinhua Li2
1Department of Thoracic and Cardiothoracic Surgery, John Muir Health.
Background:
This study aimed to quantify radiation doses during navigational bronchoscopy procedures, comparing them with reported cohorts and evaluating the LungVision (Body Vision Medical Inc.) system's efficacy in dose reduction.
Methods:
This retrospective observational study included 52 consecutive navigational bronchoscopy cases, categorized into 4 imaging groups based on the C-arm: Cios Spin (Siemens Healthineers), or OEC 9900 (GE HealthCare); and the 3D tomographic imaging algorithm: Cios Spin's onboard imaging, or LungVision's AI-driven imaging. Patient and lesion data, outcomes, and radiation indices were collected. Existing literature on 3D image guidance for bronchoscopic lung nodules was reviewed to compare reported radiation doses.
Results:
Combining LungVision with Cios Spin significantly reduced radiation dose in all cases compared with using Cios Spin alone: Cumulative air kerma (Ka,r) reduced from 238.7 to 119.1 mGy (P=0.03), and air kerma-area product (KAP) decreased from 28.19 to 15.09 Gy·cm2 (P=0.03). For biopsy cases, LungVision led to notable dose reductions: Ka,r of 279 to 129.1 mGy, and KAP of 30.70 to 16.27 Gy·cm2. LungVision notably reduced radiation indices in 7 paired spins, isolating the 3D imaging algorithm as the sole variable with the same Cios Spin C-arm. A literature review provides additional context on radiation for bronchoscopic biopsies.
Conclusion:
Following the "as low as reasonably achievable" (ALARA) principle minimizes ionizing radiation exposure, benefiting patients and operators. Physicians should compare baseline radiation levels with the literature and adopt dose-reduction techniques. LungVision's lower dose indices render it effective for real-time 3D imaging during navigational bronchoscopy while reducing radiation dose.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Radiological Investigation I: X-ray and CT