Red cell antibodies or noise? A case series on reactivity against the ingredients in column matrix
Durba Biswas1, Debapriya Basu1, Mahua Reddy1
1Department of Transfusion Medicine, Tata Medical Center, Kolkata, West Bengal, India.
Abstract:
Noise in the immunohematological investigations can be described as a false reactivity of red blood cells (RBCs) in serologic testing that is not related to the interaction of RBC antigens and antibodies that the test system is intended to detect. These false-positive reactions can cause confusion during the cross-matching and RBC antibody screening and may result in delays in patient transfusion. Although these antibodies are predominantly clinically insignificant, proper laboratory work-up is indicated to avoid misidentification of a clinically significant antibody as a noise. In this report, we describe the three rare cases where the reactivity was found against the ingredients of the column matrix (glass beads). It is imperative that such reactivity is recognized and resolved during the investigation of blood group discrepancies, positive RBC antibody screens and in cases of positive cross-matches.
Related Concept Videos
Blood Typing
Antigens are protein molecules that reside on the surface of red blood cells (RBCs). The ABO and Rh blood typing systems target...
Cross-reactivity
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Rh Blood Group
Types Of Column Chromatography
Gel Filtration Chromatography
When the...