Related Experiment Video
Updated: Jun 28, 2026

ELIME Enzyme Linked Immuno Magnetic Electrochemical Method for Mycotoxin Detection
Published on: October 23, 2009
Multiplex immunochromatographic assay using a universal labeling reactant for determining antibiotic residues in milk
Assem N Jangulova1, Nadezhda A Taranova2, Boris B Dzantiev2
1Department of Veterinary Medicine, Faculty of Veterinary and Livestock Technology, S. Seifullin Kazakh Agrotechnical Research University, 62 Zhenis Avenue, Astana 010011, Kazakhstan.
This study developed a novel multiplex immunochromatographic assay (ICA) for detecting streptomycin, tetracycline, and chloramphenicol residues in milk. The new method simplifies manufacturing by using a single reagent, reducing costs and improving antibiotic residue testing in food safety.
Area of Science:
- Food Safety and Analytical Chemistry
- Immunotechnology and Biosensing
- Veterinary Pharmacology and Residue Analysis
Background:
- Antibiotics in animal husbandry pose risks as food contaminants and drivers of antibiotic resistance.
- Regulatory bodies set strict limits for antibiotic residues in food products.
- Current immunochromatographic assays (ICAs) for antibiotic detection are complex and costly due to individual antibody labeling.
Purpose of the Study:
- To develop a cost-effective multiplex immunochromatographic assay (ICA) for simultaneous detection of streptomycin (STR), tetracycline (TC), and chloramphenicol (CAP) in milk.
- To simplify the manufacturing process of antibiotic testing strips by using indirect labeling of antibodies.
Main Methods:
- Constructed test strips using nitrocellulose membranes and various absorption/separation pads.
- Utilized specific antibiotic-conjugates (e.g., TC-BSA, STR-BSA, CAP-STI) and monoclonal antibodies (anti-TC, anti-STR, anti-CAP).
- Employed a single gold nanoparticle (GNP)-conjugated reagent (goat anti-mouse IgG) for indirect labeling of immune complexes.
Main Results:
- Successfully developed a multiplex ICA capable of detecting three antibiotic residues in milk.
- Achieved visual limits of detection (LODs) of 600 ng/mL for STR, 10 ng/mL for TC, and 30 ng/mL for CAP.
- Instrumental analysis demonstrated higher sensitivity, with LODs as low as 1.2 ng/mL (STR), 0.05 ng/mL (TC), and 1.3 ng/mL (CAP) in goat milk.
Conclusions:
- The developed multiplex ICA offers a simplified and potentially more economical approach for detecting multiple antibiotic residues in milk.
- The indirect labeling strategy using a single GNP-conjugated reagent significantly streamlines the manufacturing of ICA test strips.
- This method provides a valuable tool for rapid and efficient food safety monitoring of antibiotic contamination.