Comparing Rotational Stability of Clareon and Eyhance Toric IOLs in a Real-World Setting
Ben Connell1,2, Ben LaHood3,4
1Eye Surgery Associates, Melbourne, VIC, Australia.
Background:
The Eyhance Toric intraocular lens (IOL) builds upon the Tecnis Toric platform, initially associated with considerable post-operative rotational instability. Version 2, the Eyhance Toric IOL has been modified to enhance rotational stability. This study evaluates the post-operative rotational stability of the Eyhance Toric IOL compared to the Clareon Toric IOL, recognized for its stable performance.
Methods:
Patients undergoing cataract surgery received either the Eyhance or Clareon Toric IOLs. Placement was guided by the Barrett Toric Calculator at baseline (P0). IOL stability, uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), and refractive astigmatism were assessed at 6-24 hours (P1) and 3 weeks to 6 months (P2) post-operatively. IOL rotational measurements were recorded at each interval.
Results:
The study included 187 patients (median age: 74 for Clareon, 79 for Eyhance, p = 0.004). No significant differences were found in UDVA, CDVA, or refractive astigmatism at P2. Median rotation from P0 to P1 (3.0 vs 4.0 degrees, p = 0.091) and P0 to P2 (1.0 vs -0.5 degrees, p = 0.482) were not statistically different. However, the Clareon IOL showed less rotation between P1 and P2 (0.0 vs 1.0 degrees, p = 0.049). Absolute rotation from P0 to P1 (4.0 degrees), P1 to P2 (1.0 vs 2.0 degrees, p = 0.064), and P0 to P2 (4.0 vs 3.5 degrees, p = 0.095) were comparable.
Conclusion:
The Eyhance Toric IOL demonstrated comparable rotational stability and visual outcomes to the Clareon Toric IOL. Modifications in the Eyhance design have successfully improved its rotational stability, positioning it as a viable alternative to the Clareon Toric IOL in clinical practice.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Properties of Enantiomers and Optical Activity
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment