Related Experiment Video
Updated: Jul 17, 2026

Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
Early and midterm results of Cook Formula stent in children with right heart disease: A single-center experience
Haritha Girija1, Chinnasamy Sivaprakasam Muthukumaran1, Rajaguru Ganesan1
1Department of Pediatric Cardiology, Apollo Children's Hospital, Chennai, Tamil Nadu, India.
Insights
The Formula stent effectively treats pediatric pulmonary artery stenosis and right ventricular outflow tract obstruction, showing good outcomes with a low reintervention rate. This approach helps prevent right ventricular damage and ensures adequate lung blood flow in children.
Area of Science:
- Pediatric Cardiology
- Interventional Cardiology
- Congenital Heart Disease
Background:
- Branch pulmonary artery (PA) stenosis requires early intervention to prevent right ventricular (RV) damage and ensure lung perfusion.
- Balloon-mounted stents are valuable tools for managing RV outflow tract (RVOT) obstruction and PA stenosis.
Purpose of the Study:
- To evaluate the clinical and angiographic outcomes of using the Formula stent in children with congenital heart disease.
- To assess the efficacy of the Formula stent in treating RVOT obstruction and branch PA stenosis.
Main Methods:
- Retrospective study of 65 children undergoing stenting for RVOT obstruction and branch PA stenosis between 2018-2024.
- Utilized the Formula stent (Cook Medical) in a tertiary pediatric cardiac center.
- Analyzed clinical and angiographic data, including stent size, procedure time, and reintervention rates.
Main Results:
- 87 stents implanted in 65 patients (median age 3 years); 4 in RVOT, 83 in branch PAs.
- 19% of stents required re-dilatation at a mean of 25 months, increasing diameter and reducing RV systolic pressure.
- Successful deployment using 5Fr-8Fr sheaths with a median procedure time of 62 minutes.
Conclusions:
- The Formula stent offers comprehensive size options for pediatric PA stenosis and RVOT obstruction.
- It requires smaller sheaths, reduces procedure time, and has minimal complications.
- The stent demonstrates an acceptable rate of reinterventions, making it a viable treatment option.
Background:
Branch pulmonary artery (PA) stenosis must be addressed early to prevent right ventricular scarring and establish lung blood flow. Balloon-mounted stents are more useful in managing right ventricular outflow tract (RVOT) obstruction and PA stenosis.
Materials And Methods:
We studied the clinical and angiographic data of children with congenital heart disease who underwent stenting for RVOT obstruction and branch PA stenosis using the Formula stent (Cook Medical, Limerick, Ireland) between 2018 and 2024 in a tertiary pediatric cardiac center in southern India.
Results:
Our patient cohort included 65 patients with a median age of 3 years (range: 3 months to 17 years). The median weight of patients was 11.5 kg (4.8-60.4 kg), with 40% weighing <10 kg. Out of the 87 stents implanted, four stents were implanted in RVOT, while the rest were branch PA stenting. The median stent length was 20 mm (12-60 mm), and the median stent diameter was 10 mm (5-10 mm). All stents were deployed using 5Fr to 8Fr sheaths or guiding catheters. The median procedure time was 62 min (20-170 min), with a median fluoroscopy time of 15 min (4-110 min). A total of 17 (19%) stents in 14 patients required further dilatation at a mean interval of 25 ± 4 months from the time of stenting. Ballooning resulted in an increase in diameter from 7.8 ± 2.2 mm to 10.2 ± 1.7 mm. Right ventricular systolic pressure had reduced from 70.3 ± 8 to 40.1 ± 1 mmHg after balloon dilatation in patients with a biventricular heart.
Conclusion:
The Formula stent provides a comprehensive stent size option, requiring smaller sheaths, reduced procedural time, minimum complications, and an acceptable rate of reinterventions.
More Related Videos
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
11:20A Rodent Model of The Ross Operation: Syngeneic Pulmonary Artery Graft Implantation in A Systemic Position
Published on: April 1, 2022
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization