On the Effect of the Patient Table on Attenuation in Myocardial Perfusion Imaging SPECT
Tamino Huxohl1, Gopesh Patel2, Wolfgang Burchert2
1Institute of Radiology, Nuclear Medicine and Molecular Imaging, Heart and Diabetes Center North Rhine-Westphalia, University Hospital (Ruhr University Bochum), Medical Faculty OWL (Bielefeld University), Bad Oeynhausen, Germany. thuxohl@hdz-nrw.de.
Background:
The topic of the effect of the patient table on attenuation in myocardial perfusion imaging (MPI) SPECT is gaining new relevance due to deep learning methods. Existing studies on this effect are old, rare and only consider phantom measurements, not patient studies. This study investigates the effect of the patient table on attenuation based on the difference between reconstructions of phantom scans and polar maps of patient studies.
Methods:
Jaszczak phantom scans are acquired according to quality control and MPI procedures. An algorithm is developed to automatically remove the patient table from the CT for attenuation correction. The scans are then reconstructed with attenuation correction either with or without the patient table in the CT. The reconstructions are compared qualitatively and on the basis of their percentage difference. In addition, a small retrospective cohort of 15 patients is examined by comparing the resulting polar maps. Polar maps are compared qualitatively and based on the segment perfusion scores.
Results:
The phantom reconstructions look qualitatively similar in both the quality control and MPI procedures. The percentage difference is highest in the lower part of the phantom, but it always remains below 17.5%. Polar maps from patient studies also look qualitatively similar. Furthermore, the segment scores are not significantly different (p=0.83).
Conclusions:
The effect of the patient table on attenuation in MPI SPECT is negligible.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET