Related Experiment Video
Updated: Jun 1, 2025

Author Spotlight: Advancing Hepatic Fibrosis Diagnosis Using Magnetic Resonance Elastography and AI
Published on: July 21, 2023
Hepatoscope 2DTE's image-based quality index enhances applicability and repeatability of liver stiffness measurement
Cindy Serdjebi1, Juliette Foucher2, Adrien Besson1
1E-Scopics, Aix-en-provence, France.
Purpose:
Hepatoscope® is an ultraportable ultrasound system with 50 Hz two-dimensional transient elastography (2DTE) for liver stiffness measurement (LSM). It provides a quality index (QI) for individual stiffness values that is based on imaging features. This study evaluated the 2DTE intra- and inter-user repeatability in patients with chronic liver diseases (CLD) for novice and expert operators across various QI conditions. We compared the performances with other imaging and non-imaging elastography techniques.
Methods:
This investigation was a prospective cross-sectional single-center study. One hundred CLD patients underwent LSMs with vibration-controlled transient elastography (VCTE™), two-dimensional shear wave elastography (2DSWE) and 2DTE. Expert and novice operators each performed two consecutive exams with 2DTE, blinded to any output. Intra-class correlation coefficient for intra-, and inter-user repeatability were calculated as well as applicability for various QI conditions, and Spearman's correlations against VCTE and 2DSWE were assessed.
Results:
Computing LSM as the median of 10 stiffness values with a QI>85 % yielded the best compromise between reliability and applicability. In this setting, expert and inter-operator repeatability showed significant improvement over the condition with no QI threshold (ICC = 0.81 and 0.79, respectively) and 95 % applicability. Novice repeatability was excellent with a more restrictive QI (ICC = 0.80). LSM with 2DTE demonstrated strong and moderate correlation with VCTE and 2DSWE (r = 0.63 and 0.55, respectively).
Conclusion:
Hepatoscope 2DTE provides a promising and reliable alternative for non-invasive liver stiffness assessment in patients with CLD. The application of the image-based QI enhances LSM reliability and consistency across novice and expert operators.