Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
High resolving power electrospray drift tube ion mobility spectrometer with heated desolvation tube
Marc-Aurèle Boillat1, Peter C Hauser1
1University of Basel, Department of Chemistry, Klingelbergstrasse 80, 4056, Basel, Switzerland.
Background:
The separation performance of drift tube ion mobility spectrometers is usually relatively weak with resolving powers well below 100. Improving this aspect requires, besides the optimization of other parameters, the challenging increase of the drift voltage as deduced from fundamental equations describing the broadening of a drifting ion swarm. We recently succeeded in constructing an improved high voltage instrument equipped with an electrospray source capable of analysing liquid samples with resolving powers above 200. This instrument was further perfected with a newly developed heated desolvation tube guiding the ions to the injection gate. The electrospray desolvation process was thus enhanced while the separation region was maintained at room temperature to keep diffusional band broadening to a minimum. A further modification was also made to the detector circuitry in order to reduce its noise.
Results:
To demonstrate the performance of the instrument several challenging ion mixtures, including two herbicides, a polymer mixture, and isomers of carnitine esters were successfully separated and detected. The resolving powers, calculated from the arrival time and peak width at half height, ranged from 123 to 228. The latter value represents 96 % of the theoretically achievable resolving power at the working drift voltage of 24.1 kV.
Significance:
Stand-alone electrospray ion mobility spectrometry is not often used and has not yet been developed to its full potential. The results with improved resolving powers demonstrate its capability as a simple technique for the analysis of liquid samples, which may often serve as a lower cost alternative to chromatography or mass spectrometry.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Analyzers: Common Types
High-Resolution Mass Spectrometry (HRMS)
Capillary Electrophoresis: Instrumentation
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...