A novel pulse-current waveform circuit for low-energy consumption and low-noise transcranial magnetic stimulation
Xinhua Tan1, Ao Guo2, Jiasheng Tian2
1School of Information Science and Engineering, Yanshan University, Qinhuangdao, China.
Introduction:
Transcranial magnetic stimulation (TMS) is widely used for the noninvasive activation of neurons in the human brain. It utilizes a pulsed magnetic field to induce electric pulses that act on the central nervous system, altering the membrane potential of nerve cells in the cerebral cortex to treat certain mental diseases. However, the effectiveness of TMS can be compromised by significant heat generation and the clicking noise produced by the pulse in the TMS coil. This study proposes a novel, non-resonant, high-frequency switching design controlled by high-frequency pulse-width modulation (PWM) voltage excitation to achieve ideal pulse-current waveforms that minimize both clicking noise and heat generation from the TMS coil.
Method:
First, a particle swarm optimization algorithm was used to optimize the pulse-current waveform, minimizing both the resistance loss and clicking noise (vibration energy) generated by the TMS coils. Next, the pulse-current waveform was modeled based on the principles of programmable transcranial magnetic stimulation circuits. The relationships between the parameters of the pulse-current waveform, vibration energy, and ohmic resistance loss in the TMS coil were explored, ensuring the necessary depolarization of the nerve membrane potential. Finally, four insulated-gate bipolar transistors, controlled by a series of PWM pulse sequences, generated the desired pulse-current duration and direction in the H-bridge circuit. The duration and slope of the rising and falling segments of the current waveform were adjusted by the PWM pulse duration.
Results:
The optimized current waveform, represented by three segmented functions, reduces heat loss and noise while inducing a greater change in neural membrane potential compared with those obtained with conventional symmetric waveforms. Spectral analysis further confirmed that the noise spectrum of the optimized current waveform, particularly the peak spectrum, is significantly lower than that of the conventional triangular symmetric waveform.
Conclusion:
The study provide a method and new ideas for low energy consumption and low-noise transcranial magnetic stimulation by using TMS circuit design techniques as well as waveform optimization.
More Related Videos
07:33Measuring Contralateral Silent Period Induced by Single-Pulse Transcranial Magnetic Stimulation to Investigate M1 Corticospinal Inhibition
Published on: August 23, 2022
05:26Author Spotlight: Low-Cost Electroencephalographic Recording System Combined with a Millimeter-Sized Coil to Transcranially Stimulate the Mouse Brain In Vivo
Published on: May 26, 2023
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Half wave rectifier
Clipper Circuit
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
Clamper Circuit
Within this circuit, the diode's orientation prompts the capacitor to charge up to the level of the most negative peak of the input signal. Upon reaching this state, the diode ceases to conduct,...