Elevated estradiol in a prepubertal female: within-laboratory systematic and practical interference screening
Esra Yıldırım Demirçin1, Sevilay Sezer1, Gönül Büyükyılmaz2
1Department of Medical Biochemistry, Ministry of Health, Ankara Bilkent City Hospital, Ankara, Turkey.
Scandinavian Journal of Clinical and Laboratory Investigation
|January 23, 2025
Summary
Estradiol immunoassay interference can lead to misdiagnosis. Liquid chromatography-mass spectrometry (LC-MS/MS) offers a more accurate method for measuring estradiol, crucial for preventing inappropriate treatments, especially in pediatric cases.
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
- Laboratory Medicine
Background:
- Analytical errors in immunoassays can stem from interfering substances.
- Such interferences may lead to unnecessary medical investigations and patient anxiety.
- Estradiol measurement is particularly susceptible to interference on immunoassay platforms.
Purpose of the Study:
- To systematically screen for estradiol interference using a practical algorithm.
- To compare the performance of different immunoassay platforms and LC-MS/MS for estradiol measurement.
- To evaluate the impact of specific interference mitigation techniques.
Main Methods:
- Method comparison involving three immunoassay platforms and one LC-MS/MS.
- Serial dilution tests to assess linearity.
- Polyethylene glycol (PEG) precipitation to identify interference.
- Heterophile antibody blocking tube assay.
Main Results:
- Deviation from linearity observed during serial dilution tests.
- Polyethylene glycol (PEG) precipitation yielded a recovery rate of 63%.
- Significant discrepancies in patient estradiol values across different immunoassay platforms (e.g., Siemens: 389.02 pmol/L vs. Abbott: <36.7 pmol/L).
- LC-MS/MS provided a substantially lower and potentially more accurate estradiol value (3.33 pmol/L).
Conclusions:
- Immunoassays are prone to interferences, potentially causing overestimated estradiol results due to cross-reactivity.
- Liquid chromatography-mass spectrometry (LC-MS/MS) demonstrates greater suitability for accurate estradiol quantification.
- LC-MS/MS can aid in preventing misdiagnosis and inappropriate treatments, particularly in pediatric patients.
More Related Videos
Related Concept Videos
Hormonal Regulation
37.6K
Hormones regulate a significant portion of digestion through activation of the neuroendocrine system. The neuroendocrine system of digestion contains many different hormones all with multiple functions that are both, directly and indirectly, involved in digestion.
37.6K
Hormonal Regulation
30.4K
The renin-aldosterone system is an endocrine system which guides the renal absorption of water and electrolytes, thus managing blood pressure and osmoregulation. Activation of the system begins in the kidneys with a small cluster of cells adjacent to the afferent and efferent blood vessels of the renal corpuscle. As the nephrons are filtering blood, juxtaglomerular cells monitor blood pressure. If they detect a decrease in pressure, they release the hormone renin into the bloodstream.