Related Experiment Video
Updated: Jul 14, 2026

Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
Comparison of Temporal Artery and Rectal Temperature Measurement During Cooling and Rewarming in Neonates Treated for
Michelle LaBella1, Karen Kelly, Kristen Carlin
1Author Affiliations: Neonatal Intensive Care Unit, Seattle Children's Hospital, Seattle, WA (Mrs LaBella, Ms Kelly, Mrs Carlin, and Dr Walsh); and Seattle Children's Research Institute, Seattle, WA (Mrs Carlin and Dr Walsh).
Background:
Finding an accurate and simple method of thermometry in the neonatal intensive care unit is important. The temporal artery thermometer (TAT) has been recommended for all ages by the manufacturer; however, there is insufficient evidence for the use of TAT in infants, especially to detect hypothermia.
Purpose:
To assess the accuracy of the TAT in hypothermic neonates in comparison to a rectal thermometer.
Methods:
This study was a naturalistic, quantitative, and observational study. Temporal artery temperatures (using Exergen TemporalScanner 5000) were compared to rectal temperatures in critically ill infants in the neonatal intensive care unit undergoing therapeutic hypothermia for hypoxic ischemic encephalopathy. Temperatures were taken during a 72-hour cooling period at 33.5 °C and a 6-hour rewarming period to normothermia of 36.5 °C. Nineteen patients and 1280 temperature measurements were included in this study.
Results:
During the cooling period, TAT and rectal temperatures had a weak correlation (r = 0.34, P < .001). The correlation during the rewarming period was much stronger (r = 0.70, P < .001), indicating less variability in measures, but not agreement. On average, regardless of period, the TAT temperatures read 0.43 °C (95% confidence interval, 0.37-0.49, P < .001) warmer than the rectal temperatures. The cooling or warming period had no effect on this difference between temperatures.
Implications For Practice And Research:
This study found that temperatures obtained with a TAT are generally warmer than the accepted standard core rectal temperature in hypothermic neonates, and we do not recommend its use in critically ill neonates who require accurate temperature readings.
Related Concept Videos
Decreased Body Temperature
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...