Pulse oximetry at two sensor placement sites in conscious foals
Heini Sofia Rossi1, Anna Kristina Mykkänen2, Jouni Juho Tapio Junnila3
1Department of Equine and Small Animal Medicine, Faculty of Veterinary Medicine, University of Helsinki, Viikintie 49, 00014, Helsinki, Finland. heini.rossi@helsinki.fi.
Background:
Pulse oximetry has not been thoroughly evaluated for assessment of oxygenation in conscious foals. Compared with invasive arterial blood sampling, it is a painless and non-invasive method for real-time monitoring of blood oxygen saturation. The aim of this prospective clinical study was to evaluate the usability, validity, and reliability of pulse oximetry at two measuring sites (lip and caudal abdominal skin fold) for blood oxygen saturation measurement in conscious foals with and without respiratory compromise. Thirty-two foals under one month of age were used. Nineteen foals had normal respiratory and cardiovascular function, and 13 had pneumonia. Pulse oximetry with a transmittance sensor was performed in triplicate on each foal's lip (n = 196 measurements) and/or skin fold (n = 338 measurements), and arterial blood sample was collected. The oxygen saturation values measured by pulse oximetry from the lip and skin fold were compared with each other (n = 58 measurement pairs) and with the calculated arterial oxygen saturation based on arterial blood samples (n = 93 measurement pairs). Furthermore, repeatability of the pulse oximetry measurements was assessed.
Results:
Measured blood oxygen saturation was clearly associated with the calculated saturation, but on average (± SD) it was 1.8 (± 3.3) percentage units higher from the lip and 5.7 (± 4.3) percentage units higher from the skin fold than the calculated saturation. In concurrent lip and skin fold measurements within a foal, the skin fold measurements were 2.4 (± 2.4) percentage units higher than the lip measurements. The repeatability of three pulse oximetry saturation measurement results was moderate to good and significantly improved when the measurement furthest from the middle-measured value was excluded. The most deviating measurement was often obtained first. Pulse oximetry in general was well tolerated and easy to perform, but as expected in conscious foals, movement and contact problems generated occasional technical difficulties in some individuals.
Conclusions:
In conscious foals, pulse oximetry with a transmittance sensor attached to the lip (but not to the skin fold) is a clinically applicable and valid method for arterial blood oxygen saturation determination. Several measurements should be obtained and outliers discarded to obtain a reliable result.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse Assessment Sites
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...