Assessment of scattered and leakage radiation from ultra-portable X-ray systems in chest imaging: An independent
Leonie E Paulis1,2,3, Roald S Schnerr1,4, Jarred Halton1
1Médecins Sans Frontières, International, Amsterdam, The Netherlands.
Abstract:
Ultraportable (UP) X-ray devices are ideal to use in community-based settings, particularly for chest X-ray (CXR) screening of tuberculosis (TB). Unfortunately, there is insufficient guidance on the radiation safety of these devices. This study aims to determine the radiation dose by UP X-ray devices to both the public and radiographers compared to international dose limits. Radiation dose measurements were performed with four UP X-ray devices that met international criteria, utilizing a clinically representative CXR set-up made with a thorax phantom. Scatter and leakage radiation dose were measured at various positions surrounding the phantom and X-ray tube, respectively. These measurements were used to calculate yearly radiation doses for different scenarios based on the median of all UP X-ray devices. From the yearly scatter doses, the minimum distances from the phantom needed to stay below the international public dose limit (1 mSv/year) were calculated. This distance was longest in the direction back towards the X-ray tube and shortest to the left/right sides of the phantom, e.g., 4.5 m and 2.5 m resp. when performing 50 exams/day, at 90 kV, 2.5 mAs and source skin distance (SSD) 1 m. Additional calculations including leakage radiation were conducted at a typical radiographer position (i.e., behind the X-ray tube), with a correction factor for wearing a lead apron. At 2 m behind the X-ray tube, a radiographer wearing a lead apron could perform 106 exams/day at 2.5 mAs and 29 exams/day at 10 mAs (90 kV, SSD 1 m), while keeping his/her radiation dose below the public dose limit (1 mSv/year) and well below the radiographer dose limit (20 mSv/year). In most CXR screening scenarios, the radiation dose of UP X-ray devices can be kept below 1 mSv/year by employing basic radiation safety rules on time, distance and shielding and using appropriate CXR exposure parameters.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation