Prediction of Seven Artificial Intelligence-Based Intraocular Lens Power Calculation Formulas in Medium-Long
Wiktor Stopyra1,2, Oleksiy Voytsekhivskyy3, Andrzej Grzybowski4,5
1MW-Med Eye Centre, 31-416 Krakow, Poland.
Life (Basel, Switzerland)
|January 25, 2025
Summary
The Hill-RBF 3.0 formula demonstrated superior accuracy in intraocular lens power calculations for medium-long Caucasian eyes. Other artificial intelligence formulas also showed promising results for precise cataract surgery outcomes.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Artificial Intelligence in Healthcare
Background:
- Accurate intraocular lens (IOL) power calculation is crucial for achieving optimal visual outcomes after cataract surgery.
- Traditional formulas may have limitations in predicting refractive outcomes, especially in specific eye types.
- Advancements in artificial intelligence (AI) offer new possibilities for enhancing IOL power calculation accuracy.
Purpose of the Study:
- To evaluate and compare the accuracy of seven AI-based IOL power calculation formulas.
- To assess formula performance using metrics like root-mean-square absolute error (RMSAE), median absolute error (MedAE), and prediction error (PE) within ±0.5 D.
- To identify the most accurate AI formula for medium-long Caucasian eyes.
Main Methods:
- Retrospective review of biometric and refractive data from 214 Caucasian patients undergoing cataract surgery.
- Application of multiple AI-based IOL power calculation formulas including Hill-RBF 3.0, Ladas Super Formula AI (LSF AI), and PEARL-DGS.
- Analysis of prediction errors at various diopter thresholds (±0.25 D, ±0.50 D, ±0.75 D, ±1.00 D).
Main Results:
- The Hill-RBF 3.0 formula achieved the lowest RMSAE (0.368) and MedAE (0.200).
- Hill-RBF 3.0, LSF AI, and PEARL-DGS showed the highest percentage of eyes with PE within ±0.50 D (86.45%, 85.51%, and 85.05%, respectively).
- All evaluated AI formulas demonstrated good performance in IOL power calculation for the studied cohort.
Conclusions:
- The Hill-RBF 3.0 formula is highly accurate for IOL power calculation in medium-long Caucasian eyes.
- AI-based formulas represent a significant advancement in achieving precise refractive outcomes in cataract surgery.
- Further research may explore the efficacy of these AI formulas across diverse patient populations and eye characteristics.
Keywords:
Hill-RBF formulaartificial intelligenceintraocular lensesmedium-long eyesphacoemulsification