How to Reliably Measure Stroke Volume Index in Pulmonary Arterial Hypertension: A Comparison of Thermodilution,
Andrea Baccelli1, Deepa Gopalan2, Rachel J Davies3
1Department of Respiratory Medicine, Royal Brompton Hospital, Guy's and St Thomas' NHS Foundation Trust, London SE1 7EH, UK.
Background:
Stroke volume index (SVI) is an important prognostic parameter in pulmonary arterial hypertension (PAH). The direct Fick (DF) method represents the gold standard for measuring it. Indirect Fick (IF) and thermodilution (TD) are simpler and widely used alternatives. However, data on the accuracy of these methods in estimating SVI in PAH are scant. We aimed to compare these different invasive methods, and in a subgroup of patients, to a non-invasive method using MRI.
Methods:
We enrolled 103 PAH patients undergoing a diagnostic or follow-up right heart catheterization at our centre (mean age 56 years, 56% female). The Bland-Altman analysis was used to assess agreement between methods. Potential demographic, clinical, and hemodynamic biases were explored. The accuracy of cardiac magnetic resonance (CMR)-derived SVI was assessed in a subset of patients.
Results:
The mean bias for IF-SVI vs. DF-SVI was -5.53 mL/min/m2 with a median percentage error (PE) of 15%. The mean bias was lower, 0.09 mL/min/m2, for TD-SVI vs. DF-SVI with a median PE of 10%. Low cardiac index and severe tricuspid regurgitation (TR) were associated with a greater bias between TD and DF. CMR-SVI showed good accuracy and precision even in patients with severe TR, compared to DF.
Conclusions:
The indirect Fick is the less reliable method to assess SVI also in PAH patients. Thermodilution is a valid alternative to direct Fick, but it should be used with caution in patients with severe TR or low cardiac index. SVI measured by cardiac MRI is a promising non-invasive alternative, especially in patients with severe TR. Our observation needs to be confirmed by other series and larger studies.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
08:05Cardiac Response to β-Adrenergic Stimulation Determined by Pressure-Volume Loop Analysis
Published on: May 19, 2021
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Imaging Studies for Cardiovascular System IV: CMRI
Aortic Regurgitation II: Clinical Features and Diagnostic Tests