Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Glucometer versus analyzer: comparable results with negligible clinical risk
Alper Gümüş1, Cihan Coşkun1, Kamil Taha Uçar1
1Department of Medical Biochemistry, Health Science University, Istanbul Basaksehir Cam and Sakura City Hospital, Istanbul, Turkey.
This study evaluated Roche Accu-Chek Inform II glucometers, finding them reliable in real-world use. While most results were low-risk, overall compliance with accuracy standards needs improvement for better glucose monitoring.
Area of Science:
- Clinical Chemistry
- Medical Device Technology
- Diabetes Management
Background:
- Accurate blood glucose monitoring is crucial for diabetes management.
- Point-of-care glucose meters are widely used but require validation in real-world clinical settings.
- Assessing the reliability of widely adopted devices like the Roche Accu-Chek Inform II is essential for patient safety and effective treatment.
Purpose of the Study:
- To evaluate the real-world reliability and accuracy of Roche Accu-Chek Inform II glucometers.
- To compare glucometer results against a laboratory reference method using established guidelines.
- To identify areas for potential improvement in glucometer performance.
Main Methods:
- Retrospective analysis of 6,695 paired capillary and venous blood glucose results.
- Comparison of Roche Accu-Chek Inform II (capillary) with Roche Cobas analyzers (venous).
- Assessment of compliance using modified ISO 15197 and CLSI EP09-A3 criteria, including Passing-Bablok regression, Bland-Altman plots, and Surveillance Error Grid (SEG) analysis.
Main Results:
- Overall compliance was 81.5%, below the 94.6% acceptance criterion.
- SEG analysis indicated 90.3% of results were in the 'No-risk' zone.
- A small positive bias (0.33 mmol/L) was observed, with mean differences within permissible limits according to Bland-Altman analysis.
- Compliance varied by concentration, with highest accuracy observed between 16.65-22.20 mmol/L.
Conclusions:
- Roche Accu-Chek Inform II glucometers demonstrate real-world reliability with most results in acceptable risk categories.
- While generally reliable, the overall compliance rate indicates a need for manufacturers to explore advancements for improved accuracy.
- Further optimization could enhance the clinical utility of these point-of-care devices for diabetes management.