Related Experiment Video
Updated: May 30, 2025

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Diagnostic Capability of Pattern Electroretinogram and Three Circumpapillary Retinal Nerve Fiber Layer Thickness
Andrew Tirsi1,2, Joby Tsai3, Aaron Abizadeh2
1Department of Ophthalmology, Manhattan Eye, Ear, and Throat Hospital/Northwell Health, New York, NY, USA.
Purpose:
To assess the diagnostic capability of pattern electroretinography (PERG) and varying circumpapillary optical coherence tomography (OCT) scan diameters in glaucoma suspects (GS).
Methods:
This is a prospective, cross-sectional study. Circumpapillary retinal nerve fiber layer thickness (RNFLT) was measured using spectral domain OCT in 49 eyes from 26 patients (36 normal, 13 GS) in three circle diameters (3.5, 4.1, and 4.7 mm). PERG measurements (Magnitude [Mag], MagnitudeD [MagD], MagnitudeD/Magnitude [MagD/Mag] ratio) were used. Based on clinical examination, participants were classified as controls or GS. Independent t-test and areas under the receiver operating characteristic curve (AUC) were obtained to determine the diagnostic capability of PERG and OCT.
Results:
Independent t-test revealed significant differences between controls and GS in age, sex, central corneal thickness (CCT), all PERG parameters, and global RNFLT (gRNFLT) in three circle diameters. All PERG parameters were correlated to all RNFLT sectors (r > 0.291; p < 0.041), except in the temporal and nasal sectors of the three circle scans. Spearman rho was highest in 3.5 and lowest in 4.7 mm circle scan. AUC demonstrated 3.5 mm gRNFLT had the highest diagnostic capability (AUC = 0.877), followed by 4.1 mm gRNFLT (AUC = 0.852), and 4.7 mm gRNFLT (AUC = 0.821). MagD showed the foremost diagnostic capability (AUC = 0.81), followed by Mag (AUC = 0.799) and MagD/Mag (AUC = 0.762).
Conclusion:
Global, superior, and inferior RNFLT in 3.5 and 4.1 mm diameters, and MagD performed best in discriminating GS from controls, suggesting that a larger scan of 4.1 mm may be equally useful in glaucoma diagnosis as the conventional 3.5 mm diameter. We recommend using PERG with OCT of 3.5 or 4.1 mm diameters for glaucoma suspect diagnosis.