Solid-state esophageal pressure sensor for the estimation of pleural pressure: a bench and first-in-human validation
Julien P van Oosten1, Nico Goedendorp2, Amne Mousa3,4
1Intensive Care, Erasmus Medical Center, Dr. Molewaterplein 40, 3015 GD, Rotterdam, The Netherlands. j.vanoosten@erasmusmc.nl.
Background:
Advanced respiratory monitoring through the measurement of esophageal pressure (Pes) as a surrogate of pleural pressure helps guiding mechanical ventilation in ICU patients. Pes measurement with an esophageal balloon catheter, the current clinical reference standard, needs complex calibrations and a multitude of factors influence its reliability. Solid-state pressure sensors might be able to overcome these limitations.
Objectives:
To evaluate the accuracy of a new solid-state Pes transducer (Pessolid). We hypothesized that measurements are non-inferior to those obtained with a properly calibrated balloon catheter (Pesbal).
Methods:
Absolute and relative solid-state sensor Pes measurements were compared to a reference pressure in a 5-day bench setup, and to simultaneously placed balloon catheters in 15 spontaneously breathing healthy volunteers and in 16 mechanically ventilated ICU patients. Bland-Altman analysis was performed using mixed effects modelling with bootstrapping to estimate bias and upper and lower limits of agreement (LoA) and their confidence intervals.
Results:
Bench study: Solid-state pressure transducers had a positive bias (Psolid - Pref) of around 1 cmH2O for the absolute minimal and maximum pressures, and no bias for pressure swings. Healthy volunteers: the solid-state transducer revealed a bias (i.e., Pessolid - Pesbal) [upper LoA; lower LoA] of 1.59 [8.21; - 5.02], - 2.32 [4.27; - 8.92] and 3.91 [11.04; - 3.23] cmH2O for end-expiratory, end-inspiratory and ΔPes values, respectively. ICU patients: the solid-state transducer showed a bias (Pessolid-Pesbal) [upper LoA; lower LoA] during controlled/assisted ventilation of: - 0.15 [1.41; - 1.72]/- 0.19 [5.23; - 5.62], 0.32 [3.45; - 2.82]/- 0.54 [4.81; - 5.90] and 0.47 [3.90; - 2.96]/0.35 [4.01; - 3.31] cmH2O for end-expiratory, end-inspiratory and ΔPes values, respectively. LoA were ≤ 2cmH2O for static measurements on controlled ventilation.
Conclusions:
The novel solid-state pressure transducer showed good accuracy on the bench, in healthy volunteers and in ventilated ICU-patients. This could contribute to the implementation of Pes as advanced respiratory monitoring technique.
Trial Registration:
Clinicaltrials.gov identifier: NCT05817968 (patient study). Registered on 18 April 2023.
More Related Videos
10:28Sensitivity Enhancement of Soft Capacitive Pressure Sensors Using a Solvent Evaporation-Based Porosity Control Technique
Published on: March 24, 2023
05:57Author Spotlight: Microfluidic Channel-Based Soft Electrodes and Their Application in Capacitive Pressure Sensing
Published on: March 17, 2023
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Pressure Gauges
Measurement of Blood Pressure
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...