Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Pre-Procedural Guidelines for Assessing Blood Pressure01:10

Pre-Procedural Guidelines for Assessing Blood Pressure

520
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
520
Special considerations while measuring blood pressure01:28

Special considerations while measuring blood pressure

702
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
702
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K
Sites for measruring blood pressure01:21

Sites for measruring blood pressure

1.4K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.4K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

649
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
649
Assessing Blood pressure in the Leg01:11

Assessing Blood pressure in the Leg

1.9K
Proper measurement of leg blood pressure is a critical skill for healthcare providers, ensuring precise and reliable readings. When performed correctly, this procedure informs patient care and enhances the efficacy of interventions. The following text outlines step-by-step guidelines to measure blood pressure in the leg, providing clarity and ease of understanding for practitioners.
Preparation:
1.9K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Risk Factors for Type 2 Endoleak after Preemptive Side Branch Embolization During EVAR for Abdominal Aortic Aneurysm.

Annals of vascular surgery·2026
Same author

Impact of a Substantial Oversizing Strategy in Chimney Endovascular Aneurysm Repair on Aortic Neck Behavior and Clinical Outcomes Compared with Open Repair for Challenging Aortic Necks.

Annals of vascular surgery·2026
Same author

Clinical benefits of covering the bronchial stump with a free pericardial fat pad after thoracoscopic lobectomy.

Journal of thoracic disease·2026
Same author

miR-765 induces angiogenesis by inhibiting dipeptidyl peptidase 4 and increasing fibroblast growth factor 2.

BBA advances·2026
Same author

Early D-Dimer Screening for Pulmonary Vein Stump Thrombosis After Lung Lobectomy.

The Annals of thoracic surgery·2026
Same author

A Case of Laparoscopic-Assisted Thoracoscopic Diaphragmatic Plication for Diaphragmatic Eventration.

Asian journal of endoscopic surgery·2026

Related Experiment Video

Updated: May 30, 2025

Fabrication of Compressed Hosiery and Measurement of its Pressure Characteristic Exerted on the Lower Limbs
08:39

Fabrication of Compressed Hosiery and Measurement of its Pressure Characteristic Exerted on the Lower Limbs

Published on: May 27, 2020

2.3K

Interface Pressures Derived from a Calibrated Bandage Applied for Compression Therapy.

Kotaro Suehiro1,2, Hitoshi Sakuda2, Takasuke Harada1

  • 1Department of Surgery and Clinical Science, Yamaguchi University Graduate School of Medicine, Ube, Yamaguchi, Japan.

Annals of Vascular Diseases
|January 29, 2025
PubMed
Summary

A calibrated bandage applied to the lower leg demonstrated consistent interface pressure (IP) regardless of application method or leg circumference. This finding suggests reliable pressure distribution for compression therapy.

Keywords:
calibrated bandagecompression therapyinterface pressure

More Related Videos

Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
06:45

Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems

Published on: May 2, 2025

216
A New Hybrid Quantitative Evaluation Model for Axillary Junctional Hemorrhage in Swine
08:27

A New Hybrid Quantitative Evaluation Model for Axillary Junctional Hemorrhage in Swine

Published on: December 6, 2024

213

Related Experiment Videos

Last Updated: May 30, 2025

Fabrication of Compressed Hosiery and Measurement of its Pressure Characteristic Exerted on the Lower Limbs
08:39

Fabrication of Compressed Hosiery and Measurement of its Pressure Characteristic Exerted on the Lower Limbs

Published on: May 27, 2020

2.3K
Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems
06:45

Development of a Benchtop Model for Evaluating the Compatibility of Wound Dressing Materials with Negative Pressure Wound Therapy Systems

Published on: May 2, 2025

216
A New Hybrid Quantitative Evaluation Model for Axillary Junctional Hemorrhage in Swine
08:27

A New Hybrid Quantitative Evaluation Model for Axillary Junctional Hemorrhage in Swine

Published on: December 6, 2024

213

Area of Science:

  • Vascular Surgery
  • Biomedical Engineering
  • Orthopedics

Background:

  • Compression therapy is crucial for managing venous disorders.
  • Accurate and consistent interface pressure (IP) application is vital for effective compression therapy.
  • Calibrated bandages aim to standardize IP delivery.

Purpose of the Study:

  • To evaluate the interface pressure (IP) and its variability when using a calibrated bandage (Biflex16) on the lower leg.
  • To determine if leg circumference influences the applied IP.
  • To assess the reliability of a specific bandage application technique.

Main Methods:

  • Two studies were conducted involving participants applying a calibrated bandage to a lower leg model or their own leg.
  • Study I compared two application methods: one with intentional stretch (Application 1) and a control (Application 2).
  • Study II utilized Application 1, measuring IP at a specific anatomical landmark (B1) in the sitting position.

Main Results:

  • In Study I, both application methods yielded similar median interface pressures (37 mmHg) and interquartile ranges (9 mmHg) in the standing position.
  • In Study II, Application 1 resulted in a median IP of 38 mmHg and an IQR of 12 mmHg in the sitting position.
  • No significant correlation was observed between the applied IP and the leg circumference.

Conclusions:

  • The calibrated bandage, when applied using Application 1, demonstrated reasonably small variations in interface pressure.
  • The applied interface pressure was consistent and did not correlate with leg circumference.
  • These findings support the potential for reliable and predictable compression therapy using this calibrated bandage system.