Observation of Ɛ-Al13-Keggin-oligonucleotide complexes in analytical sample solutions
Jonathan C Chan1, Bao V Mai1, Andrew A Rodriguez1
1Ionis Pharmaceuticals, Inc., 2855 Gazelle Ct., Carlsbad, CA 92010, United States.
Journal of Pharmaceutical Sciences
|January 29, 2025
Summary
Aluminum cluster molecules (Ɛ-Al13-Keggin) can form artifact impurities with antisense oligonucleotides during analysis. Using alternative materials like polypropylene or glass for sample preparation prevents this unwanted complex formation.
Area of Science:
- Analytical Chemistry
- Materials Science
- Biochemistry
Background:
- Drug product stability testing is crucial for pharmaceutical quality control.
- Antisense oligonucleotides are a class of therapeutic nucleic acids.
- Aluminum cluster molecules, such as Ɛ-Al13-Keggin, can be encountered in analytical settings.
Purpose of the Study:
- To identify the source of new impurity peaks observed during antisense oligonucleotide drug product stability testing.
- To characterize the interaction between Ɛ-Al13-Keggin molecules and antisense oligonucleotides.
- To develop strategies to prevent artifact formation during sample preparation.
Main Methods:
- Impurity profiling using analytical techniques.
- Synthesis and characterization of Ɛ-Al13-Keggin molecules.
- Binding affinity studies between Ɛ-Al13-Keggin and various oligonucleotide sequences.
- Comparative analysis of sample preparation methods using different weighing materials.
Main Results:
- Observed impurity peaks were identified as complexes of Ɛ-Al13-Keggin molecules and antisense oligonucleotides.
- These complexes were determined to be analytical artifacts arising from the use of aluminum weigh boats.
- Ɛ-Al13-Keggin molecules demonstrated binding affinity to oligonucleotides of varying lengths and compositions, with a potential preference for 5-methylcytosine.
- Switching to polypropylene or glass weighing funnels eliminated the formation of these artifact complexes.
Conclusions:
- The use of aluminum weigh boats in sample preparation can lead to artifactual complex formation between Ɛ-Al13-Keggin molecules and antisense oligonucleotides.
- Modification of sample preparation procedures is essential to ensure accurate stability testing of oligonucleotide drug products.
- Alternative inert materials for sample handling prevent the co-precipitation of analytical artifacts.