Related Experiment Video
Updated: May 30, 2025

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Case report: Clinical workflow considerations for treating soft-tissue sarcoma on a 1.5-T MR-Linac
Gye Won Choi1, Gary A Eastwick1, Leonard H Kim1
1MD Anderson Cancer Center at Cooper, 2 Cooper Plaza, Camden, NJ 08103, USA.
Abstract:
We present specific issues that arose when using a 1.5-Tesla MR-Linac to treat a series of 4 soft-tissue sarcoma (STS) patients. These issues arose from the combination of typical STS attributes (long, off-axis target) and MR-Linac design-specific limitations on field size and patient positioning. Despite the availability of on-line plan adaptation, STS patients were more efficiently treated after workflow changes to improve patient selection and immobilization. Other issues arising from off-axis STS target locations: geometric distortion of MR images and patient-specific QA, are discussed.