Accuracy of a Novel High-Throughput "Car Blood Pressure" Measurement Protocol
Neiberg de Alcantara Lima1,2, Shaun Cardozo1, Andrew Johnson3
1Division of Cardiovascular Diseases, Department of Internal Medicine, Wayne State University, Detroit, MI, USA.
Background:
More than one in three adults with hypertension in the United States are unaware of their condition, highlighting the importance of large-scale screening campaigns. Currently, the identification of hypertension is largely limited to medical settings. To help overcome this barrier, we developed a novel high-throughput screening protocol that measures blood pressure (BP) while patients remain seated in an automobile ("car-BP"). The aim of this study was to provide an initial assessment of the accuracy of car-BP.
Methods:
Three BP readings were determined in a clinic exam room before and after three BP readings were taken while patients were seated in a parked car outside (n = 100 participants). The same validated device model (Omron HEM-907XL) and BP measurement methods adhering to guidelines were used in both scenarios. The average of all 6 clinic readings was compared to the average of the 3 car-BP readings in each individual.
Results:
Mean clinic and car-BP readings were 120.9 ± 16.2/78.0 ± 9.9 and 118.9 ± 15.2/76.0 ± 10.0 mm Hg, respectively. The paired mean and absolute mean differences in systolic BP levels between methods were -1.92 mm Hg (95% confidence interval (CI) -3.2 to -0.7 mm Hg) and 4.8 mm Hg (95%CI 3.8 to 5.6 mm Hg), respectively. A total of 85% of participants had both systolic and diastolic BP levels ≤ 10 mm Hg different between measurement scenarios (meeting the a priori determined study primary outcome).
Conclusions:
Car-BP represents an innovative and accessible approach for potential large-scale hypertension screening campaigns.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement